APPLIED BIOSYSTEMS 7500 FAST DX REAL TIME PCR
Report
- Report Number
- 3003673482-2014-00019
- Event Type
- Malfunction
- Date Received
- June 27, 2014
- Date of Event
- June 22, 2013
- Report Date
- June 22, 2013
- Manufacturer
- LIFE TECHNOLOGIES HOLDINGS PTE LTD
- Product Code
- NSU
- PMA / PMN Number
- K082562
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. REPAIR WAS DONE BY REPLACING LAMP, LAMP HOLDER AND LAMP CABLE. THIS DEVICE IS USED TO OBTAIN RESULTS FOR A DIAGNOSIS. DEVICE INTENDED USE: THE APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT WITH THE SDS SOFTWARE VERSION 1.4 IS A REAL-TIME NUCLEIC ACID AMPLIFICATION AND DETECTION SYSTEM THAT MEASURES NUCLEIC ACID SIGNALS FROM REVERSE TRANSCRIBED RNA AND CONVERTS THEM TO COMPARATIVE QUANTITATIVE READOUTS USING FLUORESCENT DETECTION OF DUAL-LABELED HYDROLYSIS PROBES. THE 7500 FAST DX REAL-TIME PCR INSTRUMENT IS TO BE USED ONLY BY TECHNOLOGISTS TRAINED IN LABORATORY TECHNIQUES, PROCEDURES AND ON USE OF THE ANALYZER.
A CUSTOMER REPORTED THAT A FLUCTUATING LAMP SIGNAL ON THE INSTRUMENT APPLIED BIOSYSTEMS 7500 FAST DX (CATALOG NO. 4407205 WITH SERIAL NO. (B)(4) PRODUCED INCONSISTENT DATA OBSERVED THROUGH INTERMITTENT BASELINE FLUCTUATIONS. NO PT INVOLVEMENT REPORTED. COMPANY EVENT TRACKING NO: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 378411 | APPLIED BIOSYSTEMS 7500 FAST DX REAL TIME PCR | INSTRUMENTATION FOR CLINICAL MULTIPL | NSU | LIFE TECHNOLOGIES HOLDINGS PTE LTD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |