APPLIED BIOSYSTEMS 7500 FAST DX REAL TIME PCR

MDR report key: 3978866 · Received June 27, 2014

Report

Report Number
3003673482-2014-00019
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
June 22, 2013
Report Date
June 22, 2013
Manufacturer
LIFE TECHNOLOGIES HOLDINGS PTE LTD
Product Code
NSU
PMA / PMN Number
K082562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. REPAIR WAS DONE BY REPLACING LAMP, LAMP HOLDER AND LAMP CABLE. THIS DEVICE IS USED TO OBTAIN RESULTS FOR A DIAGNOSIS. DEVICE INTENDED USE: THE APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT WITH THE SDS SOFTWARE VERSION 1.4 IS A REAL-TIME NUCLEIC ACID AMPLIFICATION AND DETECTION SYSTEM THAT MEASURES NUCLEIC ACID SIGNALS FROM REVERSE TRANSCRIBED RNA AND CONVERTS THEM TO COMPARATIVE QUANTITATIVE READOUTS USING FLUORESCENT DETECTION OF DUAL-LABELED HYDROLYSIS PROBES. THE 7500 FAST DX REAL-TIME PCR INSTRUMENT IS TO BE USED ONLY BY TECHNOLOGISTS TRAINED IN LABORATORY TECHNIQUES, PROCEDURES AND ON USE OF THE ANALYZER.

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT A FLUCTUATING LAMP SIGNAL ON THE INSTRUMENT APPLIED BIOSYSTEMS 7500 FAST DX (CATALOG NO. 4407205 WITH SERIAL NO. (B)(4) PRODUCED INCONSISTENT DATA OBSERVED THROUGH INTERMITTENT BASELINE FLUCTUATIONS. NO PT INVOLVEMENT REPORTED. COMPANY EVENT TRACKING NO: (B)(4).