APPLIED BIOSYSTEMS 7500 FAST DX REAL TIME PCR
Report
- Report Number
- 3003673482-2014-00021
- Event Type
- Malfunction
- Date Received
- June 27, 2014
- Date of Event
- January 16, 2014
- Report Date
- January 16, 2014
- Manufacturer
- LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
- Product Code
- NSU
- PMA / PMN Number
- K082562
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE REPAIR COMPLETED BY REPLACING HEATER COVER. THE INSTRUMENT PASSED ALL SPECIFICATION TESTS AFTER THE HEATER COVERED WAS REPLACED. THIS DEVICE IS USED TO OBTAIN RESULTS FOR A DIAGNOSIS. DEVICE INTENDED USE: THE APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT WITH THE SDS SOFTWARE VERSION 1.4 IS A REAL-TIME NUCLEIC ACID AMPLIFICATION AND DETECTION SYSTEM THAT MEASURES NUCLEIC ACID SIGNALS FROM REVERSE TRANSCRIBED RNA AND CONVERTS THE TO COMPARATIVE QUANTITATIVE READOUTS USING FLUORESCENT DETECTION OF DUAL-LABELED HYDROLYSIS PROBES. THE 7500 FAST DX REAL-TIME PCR INSTRUMENT IS TO BE USED ONLY BY TECHNOLOGISTS TRAINED IN LABORATORY TECHNIQUES, PROCEDURES, AND ON USE OF THE ANALYZER.
A CUSTOMER REPORTED CONTAMINATION ON HEATER COVER ON THE INSTRUMENT APPLIED BIOSYSTEMS 7500 FAST DX (CAT. NO. 4406985) WITH SERIAL NO. (B)(4), RESULTING IN INCORRECT RESULTS REPORTED FOR AN ASSAY RUN. NO PT INVOLVEMENT REPORTED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 378387 | APPLIED BIOSYSTEMS 7500 FAST DX REAL TIME PCR | INSTRUMENTATION FOR CLINICAL MULTIPLEX | NSU | LIFE TECHNOLOGIES HOLDINGS PTE. LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |