FDA Adverse Event Malfunction Summary report: N

APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR

MDR report key: 3978862 · Received June 27, 2014

Report

Report Number
3003673482-2014-00023
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
September 19, 2012
Report Date
September 19, 2012
Manufacturer
LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
Product Code
NSU
PMA / PMN Number
K082562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE FIELD SERVICE ENGINEER (FSE) REVIEWED DATA FILES AND REPLACED THE POWER AMPLIFIER TO RESOLVE THE ISSUE. DEVICE INTENDED USE: THE APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT WITH THE SDS SOFTWARE VERSION 1.4 IS AN REAL-TIME NUCLEIC ACID AMPLIFICATION AND DETECTION SYSTEM THAT MEASURES NUCLEIC ACID SIGNALS FROM REVERSE TRANSCRIBED RNA AND CONVERTS THEM TO COMPARATIVE QUANTITATIVE READOUTS USING FLUORESCENT DETECTION OF DUAL-LABELED HYDROLYSIS PROBES. THE 7500 FAST DX REAL-TIME PCR INSTRUMENT IS TO BE USED ONLY BY TECHNOLOGISTS TRAINED IN LABORATORY TECHNIQUES, PROCEDURES, AND ON USE OF THE ANALYZER.

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT (CATALOG NO. 4407205 SERIAL NO. (B)(4)) RESET ITSELF IN THE MIDDLE OF A RUN. WE WERE NOT MADE AWARE OF ANY IMPACT TO PATIENT MANAGEMENT. THE COMPANY EVENT TRACKING NO. IS (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
377629 APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENTATION FOR CLINICAL MULTIP NSU LIFE TECHNOLOGIES HOLDINGS PTE. LTD.

Patients

Seq Age Sex Outcome Treatment
1