FDA Adverse Event Malfunction Summary report: N

APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR

MDR report key: 3978858 · Received June 27, 2014

Report

Report Number
3003673482-2014-00026
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
April 12, 2012
Report Date
April 12, 2012
Manufacturer
LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
Product Code
NSU
PMA / PMN Number
K082562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE FIELD SERVICE ENGINEER VISUALLY INSPECTED THE DEVICE AND FOUND THAT THE CONNECTOR AND CABLE PLUG HAD BURN MARKS. THE FSE REPLACED THE LAMP CABLE AND CONTROLLER BOARD. DEVICE INTENDED USE: THE APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT WITH THE SDS SOFTWARE VERSION 1.4 IS AN REAL-TIME NUCLEIC ACID AMPLIFICATION AND DETECTION SYSTEM THAT MEASURES NUCLEIC ACID SIGNALS FROM REVERSE TRANSCRIBED RNA AND CONVERTS THEM TO COMPARATIVE QUANTITATIVE READOUTS USING FLUORESCENT DETECTION OF DUAL-LABELED HYDROLYSIS PROBES. THE 7500 FAST DX REAL-TIME PCR INSTRUMENT IS TO BE USED ONLY BY TECHNOLOGISTS TRAINED IN LABORATORY TECHNIQUES, PROCEDURES, AND ON USE OF THE ANALYZER.

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT (CATALOG NO. 4407205 SERIAL NO. (B)(4) SHOWS HIGH BACKGROUND IN THE AMPLIFICATION PLOT RESULTS. WE WERE NOT MADE AWARE OF ANY IMPACT TO PT MANAGEMENT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
377255 APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTTRUMENTATION FOR CLINICAL MULTIPL NSU LIFE TECHNOLOGIES HOLDINGS PTE. LTD.

Patients

Seq Age Sex Outcome Treatment
1