APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR
Report
- Report Number
- 3003673482-2014-00030
- Event Type
- Malfunction
- Date Received
- June 27, 2014
- Date of Event
- January 22, 2014
- Report Date
- January 22, 2014
- Manufacturer
- LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
- Product Code
- NSU
- PMA / PMN Number
- K082562
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE INSTRUMENT WAS INVESTIGATED ON SITE. IT WAS REPAIRED ON SITE BY REPLACING THE LAMP HOUSING AND THE THICKER LAMP POWER CABLE. DEVICE INTENDED USE: THE APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT WITH THE SDS SOFTWARE VERSION 1.4 IS AN REAL-TIME NUCLEIC ACID AMPLIFICATION AND DETECTION SYSTEM THAT MEASURES NUCLEIC ACID SIGNALS FROM REVERSE TRANSCRIBED RNA AND CONVERTS THEM TO COMPARATIVE QUANTITATIVE READOUTS USING FLUORESCENT DETECTION OF DUAL-LABELED HYDROLYSIS PROBES. THE 7500 FAST DX REAL-TIME PCR INSTRUMENT IS TO BE USED ONLY BY TECHNOLOGISTS TRAINED IN LABORATORY TECHNIQUES, PROCEDURES, AND ON USE OF THE ANALYZER.
A CUSTOMER REPORTED SPIKES IN THE AMPLIFICATION PLOT OUTPUTS FROM THE INSTRUMENT, APPLIED BIOSYSTEMS 7500 FAST DX (CAT. NO. 4406985, SERIAL NO. (B)(4)). NO PATIENT INVOLVEMENT REPORTED. THE COMPANY EVENT TRACKING NO. IS (B)(4) .
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 377251 | APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR | INSTRUMENTATION FOR CLINICAL MULTIP | NSU | LIFE TECHNOLOGIES HOLDINGS PTE. LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |