FDA Adverse Event Malfunction Summary report: N

APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR

MDR report key: 3978837 · Received June 27, 2014

Report

Report Number
3003673482-2014-00030
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
January 22, 2014
Report Date
January 22, 2014
Manufacturer
LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
Product Code
NSU
PMA / PMN Number
K082562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE INSTRUMENT WAS INVESTIGATED ON SITE. IT WAS REPAIRED ON SITE BY REPLACING THE LAMP HOUSING AND THE THICKER LAMP POWER CABLE. DEVICE INTENDED USE: THE APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT WITH THE SDS SOFTWARE VERSION 1.4 IS AN REAL-TIME NUCLEIC ACID AMPLIFICATION AND DETECTION SYSTEM THAT MEASURES NUCLEIC ACID SIGNALS FROM REVERSE TRANSCRIBED RNA AND CONVERTS THEM TO COMPARATIVE QUANTITATIVE READOUTS USING FLUORESCENT DETECTION OF DUAL-LABELED HYDROLYSIS PROBES. THE 7500 FAST DX REAL-TIME PCR INSTRUMENT IS TO BE USED ONLY BY TECHNOLOGISTS TRAINED IN LABORATORY TECHNIQUES, PROCEDURES, AND ON USE OF THE ANALYZER.

Description of Event or Problem · 1

A CUSTOMER REPORTED SPIKES IN THE AMPLIFICATION PLOT OUTPUTS FROM THE INSTRUMENT, APPLIED BIOSYSTEMS 7500 FAST DX (CAT. NO. 4406985, SERIAL NO. (B)(4)). NO PATIENT INVOLVEMENT REPORTED. THE COMPANY EVENT TRACKING NO. IS (B)(4) .

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
377251 APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENTATION FOR CLINICAL MULTIP NSU LIFE TECHNOLOGIES HOLDINGS PTE. LTD.

Patients

Seq Age Sex Outcome Treatment
1