MICROFRANCE® INSTRUMENT
Report
- Report Number
- 9680837-2014-00064
- Event Type
- Malfunction
- Date Received
- August 4, 2014
- Date of Event
- July 10, 2014
- Report Date
- July 10, 2014
- Manufacturer
- XOMED MICROFRANCE MFG
- Product Code
- GEI
- PMA / PMN Number
- K993655
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES. CONCOMITANT MEDICAL PRODUCTS: - CEV634-1A (QTY. 2) - BIPOLAR INSERT CEV634-1A MOUIEL, LOT NUMBER 140414. - CEV669B - HANDLE CEV669B DIA 5MM ANG BIPOLAR, LOT NUMBER 110606. (B)(4). PRODUCT EVALUATION: ANALYSIS ON THE BIPOLAR INSERTS (CEV634-1A QTY. 3), FOUND THAT THE INSERTS ARE IN SHORT CIRCUIT. THEY WERE NOT USED AND SO WERE NOT OF ANY DANGER FOR THE PATIENT OR THE USER. AFTER DISASSEMBLY OF THE WIRES, SOME IMPACTS ARE OBSERVABLE ON THE COATING. THIS PART IS NOT IN PATIENT CONTACT, THE RISK WOULD HAVE BEEN VERY LOW IF THE INSERTS HAVE BEEN USED. THE EXACT CAUSE CANNOT BE DETERMINED WITH ABSOLUTE CERTAINTY. HOWEVER, IT WOULD BE UNLIKELY TO HAVE SOME NEW INSERTS WITH DIFFERENT LOTS, ALL IN SHORT CIRCUIT BECAUSE OF A MANUFACTURING DEFECT. THE TESTS PERFORMED BY THE CUSTOMER MAY HAVE DETERIORATED THE COATING OF THE INSTRUMENTS, WHICH LED TO THE SHORT CIRCUITS. THE DEVICES FAILED ELECTRICAL TESTING. ANALYSIS ON THE HANDLE (CEV669B) FOUND THAT THE BLACK PLASTIC PART IS BURNT NEXT TO THE BIPOLAR CONNECTORS. THE BURN OF THE PLASTIC PART IS THE CONSEQUENCE OF THE FORMATION OF AN ELECTRIC ARC, PROBABLY DUE TO THE PRESENCE OF HUMIDITY IN THE CONNECTION ZONE (CABLE NOT DRIED / BLOOD / TISSUES). THE DEVICE FAILED ELECTRICAL TESTING.
IT WAS REPORTED THAT WHEN CHECKING ¿OUR BIPOLAR ELECTRODES WITH TESTING EQUIPMENT, SOME ELECTRICAL ARCS APPEARS BETWEEN THE TWO CONNECTORS. ¿IF WE ASSEMBLE THE ELECTRODE WITH A CEV679B TUBE AND A CEV669B HANDLE, SOME SPARKS APPEAR BETWEEN THE CONNECTORS OF THE HANDLE¿ IF WE PLUG A CABLE ON THE HANDLE, WE DO NOT SEE THE SPARKS ANYMORE BUT WE CAN HEAR A SIGNIFICANT CRACKLING¿¿¿ THERE WAS NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 454337 | MICROFRANCE® INSTRUMENT | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | XOMED MICROFRANCE MFG | CEV634-1A | 140409 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |