FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 3974909 · Received August 4, 2014

Report

Report Number
1031452-2014-05141
Event Type
Malfunction
Date Received
August 4, 2014
Report Date
June 30, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

(B)(4). NO INJURY ALLEGED. MALFUNCTION ALLEGED - PROVIDER STATES THE PILOT VALVE IS NOT SWITCHING. PER REPAIR STATEMENT THE UNIT HAS LOW 02 OR YELLOW LIGHT, HIGH PRESSURE/ NOT SWITCHING, COMPRESSOR / HEAT EXCHANGER, CLAMP ON THE MANIFOLD IS LEAKING, O-RING ON THE MANIFOLD HAS BEEN REPLACED, MANIFOLD NOT SWITCHING, REX ROTH ON THE MANIFOLD IS LEAKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
453285 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other