FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3974842 · Received June 11, 2014

Report

Report Number
3008642652-2014-01754
Event Type
Malfunction
Date Received
June 11, 2014
Date of Event
May 14, 2014
Report Date
June 5, 2014
Manufacturer
ZOLL LIFE CORPORATION
Product Code
MVK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) WAS COMPLETED. UPON INVESTIGATION THE CABLE CONNECTING THE DISTRIBUTION NODE TO THE REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF, DAMAGING WIRES. THE ROOT CAUSE FOR THE STRAINED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE LAST PT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, ELECTRODE BELT SN (B)(4) FAILED INCOMING FUNCTIONALITY TESTING. THE LAST PT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
343810 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFE CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA