FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 3974740
·
Received August 4, 2014
Report
- Report Number
- 1031452-2014-04927
- Event Type
- Malfunction
- Date Received
- August 4, 2014
- Report Date
- June 20, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
DURING TESTING AND REPAIR AT THE SERVICE CENTER, THE DEVICE ALARMED FOR LOW O2. THIS WAS DUE TO A LOOSE MANIFOLD GEAR CLAMP.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE GEAR CLAMP IS LOOSE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 454210 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC5PO2V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |