FDA Adverse Event Malfunction Summary report: N

UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM

MDR report key: 3974724 · Received August 4, 2014

Report

Report Number
1061932-2014-01788
Event Type
Malfunction
Date Received
August 4, 2014
Date of Event
July 11, 2014
Report Date
July 11, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K081930
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(6). A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE. THE FSE EVALUATED THE INSTRUMENT AND DISCOVERED A STICKY SOLENOID VL230 WHICH WAS CAUSING THE LEAK. THE FSE PROCEEDED TO REPLACE VL230 TO RESOLVE THE LEAK. THE FSE VERIFIED THE REPAIR AS PER ESTABLISHED PROCEDURES AND RESULTS MET PUBLISHED SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A LEAK FROM THE FRONT OF THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM. THE CUSTOMER STATED THAT APPROXIMATELY ONE HUNDRED (100) MILLILITERS OF FLUID LEAKED AND WAS CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF GLOVES AND A LABORATORY COAT AT THE TIME OF THE EVENT AND THERE WAS NO REPORT OF INJURY OR DIRECT EXPOSURE TO THE LEAK. ERRONEOUS RESULTS WERE NOT GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
454980 UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1