FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 3974708 · Received August 3, 2014

Report

Report Number
2955842-2014-04677
Event Type
Malfunction
Date Received
August 3, 2014
Date of Event
June 23, 2014
Report Date
July 8, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THAT THE PITCH DOWN CABLE WAS FRAYED AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE FRAYED CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SURGICAL PROCEDURE, THE PROGRASP FORCEPS INSTRUMENT WAS OBSERVED TO BE BROKEN AT THE FRONT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
452619 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-10 M11131129 916

Patients

Seq Age Sex Outcome Treatment
1 22 YR