ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-22158
- Event Type
- Malfunction
- Date Received
- August 3, 2014
- Report Date
- July 21, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 10/07/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP WAS EXERCISED FOR 24 HOURS WITH THE USERS PROGRAMMED BASAL AND I:C RATIO SETTINGS OBTAINED FROM THE PUMP DOWNLOAD. AFTER 24 HOURS, THE PUMP WAS TURNED OFF AND THE BATTERY WAS REMOVED AND THEN REINSERTED IN ORDER TO STIMULATE A BATTERY REPLACEMENT. THE BASAL SETTINGS AND I:C RATIO REMAINED CORRECTLY PROGRAMMED IN THE PUMP SETTINGS WITH NO CHANGES OBSERVED. THE INITIAL COMPLAINT COULD NOT BE DUPLICATED. UNRELATED TO THE INITIAL COMPLAINT, THE BATTERY COMPARTMENT WAS CRACKED BETWEEN THE BUMPER PAD AND THE COVER. THE RETURNED BATTERY CAP WAS ABLE TO FULLY ATTACH TO THE PUMP AND WAS USED TO COMPLETE ALL TESTING.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE BASAL AND INSULIN TO CARBOHYDRATE RATIO SETTINGS WERE RESET AFTER A BATTERY CHANGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 452536 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |