FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 3974647 · Received August 3, 2014

Report

Report Number
3004209178-2014-89229
Event Type
Malfunction
Date Received
August 3, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM. INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE KEYPAD TRACES. MINOR SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER'S MOTHER REPORTED RECEIVING A BUTTON ERROR ALARM ON THE INSULIN PUMP. CUSTOMER'S CURRENT BLOOD GLUCOSE READING IS 157 MG/DL. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
452401 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAB

Patients

Seq Age Sex Outcome Treatment
1 8 YR