FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3974634 · Received August 3, 2014

Report

Report Number
3004209178-2014-89230
Event Type
Malfunction
Date Received
August 3, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A CRACKED RESERVOIR TUBE LIP WAS NOTED DURING THE VISUAL INSPECTION. THE INSULIN PUMP ALARMED DURING THE REWIND DUE TO A CORRODED MOTOR HOME SWITCH. THE MOTOR ERROR ALARM COULD NOT BE CONFIRMED DUE TO THIS ALARM. HOWEVER, THE INSULIN PUMP PASSED THE DISPLACEMENT TEST.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING A MOTOR ERROR ALARM ON THE INSULIN PUMP. CUSTOMER STATED THAT THE INSULIN PUMP WOULD NOT REWIND. CUSTOMER'S CURRENT BLOOD GLUCOSE READING IS 347 MG/DL. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
452379 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522LNAB

Patients

Seq Age Sex Outcome Treatment
1 59 YR