FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3974634
·
Received August 3, 2014
Report
- Report Number
- 3004209178-2014-89230
- Event Type
- Malfunction
- Date Received
- August 3, 2014
- Date of Event
- July 2, 2014
- Report Date
- July 2, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A CRACKED RESERVOIR TUBE LIP WAS NOTED DURING THE VISUAL INSPECTION. THE INSULIN PUMP ALARMED DURING THE REWIND DUE TO A CORRODED MOTOR HOME SWITCH. THE MOTOR ERROR ALARM COULD NOT BE CONFIRMED DUE TO THIS ALARM. HOWEVER, THE INSULIN PUMP PASSED THE DISPLACEMENT TEST.
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
CUSTOMER REPORTED RECEIVING A MOTOR ERROR ALARM ON THE INSULIN PUMP. CUSTOMER STATED THAT THE INSULIN PUMP WOULD NOT REWIND. CUSTOMER'S CURRENT BLOOD GLUCOSE READING IS 347 MG/DL. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 452379 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-522LNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |