FDA Adverse Event Malfunction Summary report: N

UNICEL DXC 800 PRO SYNCHRON SYSTEM

MDR report key: 3974363 · Received August 1, 2014

Report

Report Number
2050012-2014-00362
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 10, 2014
Report Date
July 10, 2014
Manufacturer
BECKMAN COULTER
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) WENT ONSITE AND FOUND THAT THE WIRING HARNESS WAS BURNT. THE FSE REPLACED THE DAMAGED WIRING HARNESS AND THE LINE CONDITIONER TO RESOLVE THE ISSUE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A CRACKLING SOUND AND BURNING SMELL COMING FROM THE UNICEL DXC 800 PRO SYNCHRON SYSTEM. THERE WAS NO REPORT OF INJURY OR ERRONEOUS RESULTS GENERATED. THERE WAS NO VISIBLE SMOKE, SPARKS OR FLAMES. THE FIRE DEPARTMENT WAS NOT CALLED. THE LABORATORY HAS A RISK MANAGEMENT PLAN IN PLACE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
450280 UNICEL DXC 800 PRO SYNCHRON SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1