FDA Adverse Event
Malfunction
Summary report: N
UNICEL DXC 800 PRO SYNCHRON SYSTEM
MDR report key: 3974363
·
Received August 1, 2014
Report
- Report Number
- 2050012-2014-00362
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- July 10, 2014
- Report Date
- July 10, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
Additional Manufacturer Narrative · 1
THE FIELD SERVICE ENGINEER (FSE) WENT ONSITE AND FOUND THAT THE WIRING HARNESS WAS BURNT. THE FSE REPLACED THE DAMAGED WIRING HARNESS AND THE LINE CONDITIONER TO RESOLVE THE ISSUE. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED A CRACKLING SOUND AND BURNING SMELL COMING FROM THE UNICEL DXC 800 PRO SYNCHRON SYSTEM. THERE WAS NO REPORT OF INJURY OR ERRONEOUS RESULTS GENERATED. THERE WAS NO VISIBLE SMOKE, SPARKS OR FLAMES. THE FIRE DEPARTMENT WAS NOT CALLED. THE LABORATORY HAS A RISK MANAGEMENT PLAN IN PLACE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 450280 | UNICEL DXC 800 PRO SYNCHRON SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |