FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3974292 · Received August 1, 2014

Report

Report Number
3004209178-2014-89139
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE

Additional Manufacturer Narrative · 1

THE INSULIN PUMP BUTTONS FUNCTIONED PROPERLY DURING THE TESTING. NO KEYPAD CONNECTOR WAS UNLOCKED. NO CORROSION OR WATER DAMAGE WAS NOTED ON THE KEYPAD TRACES. THE INSULIN PUMP HAD A CRACKED CASE ON THE DISPLAY WINDOW CORNERS, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP'S DOWN ARROW BUTTON WAS NOT RESPONSIVE. CUSTOMER STATED THEY REMOVED AND REINSERTED THE BATTERIES, HOWEVER THE ISSUE RETURNED. CUSTOMER STATED THEY DO NOT RECALL ANY SIGNIFICANT EVENTS THAT LEAD TO THE ALARM OR IF THE DEVICE WAS DROPPED. ADVISED CUSTOMER TO DISCONTINUE USE OF DEVICE AND REVERT TO BACK UP PLAN. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 196 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
449992 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 21 YR