FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVERINSTRUMENT

MDR report key: 3974255 · Received August 1, 2014

Report

Report Number
2955842-2014-04668
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 8, 2014
Report Date
July 8, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS FOUND ONE GRIP CABLE FRAYED AT THE DISTAL CLEVIS HUB. FRAYED STRANDS STUCK OUT APPROXIMATELY .215. THE SAME GRIP CABLE WAS ALSO DERAILED AT THE DISTAL IDLER AS WELL AS THE CLAMPING PULLEY AT THE PROXIMAL END. AS A RESULT THE INSTRUMENT WAS STUCK IN A YAW POSITION. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE FRAYED CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SURGICAL PROCEDURE, A CABLE WAS PROTRUDING FROM THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT. THE INSTRUMENT WAS NOT USED ON THE PATIENT AFTER THE REPORTED ISSUE WAS IDENTIFIED AND THERE WAS NO ALLEGATION OF HARM OR INJURY OR REPORTS OF FRAGMENTS FALLING INTO A PATIENT.