FDA Adverse Event Malfunction Summary report: N

5MM PEEK MULTIFUNCTION HANDLE

MDR report key: 3974169 · Received August 1, 2014

Report

Report Number
0002936485-2014-00558
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 9, 2014
Report Date
July 9, 2014
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
KOG
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED FOR INSLUATION FAILURE BUT NOT THE HANDLE. DENTS AND SCRATCHES WERE SEEN ON THE INSULATION. THE INSULATION WAS TESTED FOR CRACKED/DAMAGES. THE INSULATION PASSED THE INSULATION SCAN TEST. THE PROBABLE ROOT CAUSES ARE USER MISUSE, IMPROPER STERILIZATION METHODS, OR NORMAL WEAR. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED FOR INSLUATION FAILURE, BUT NOT THE HANDLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
451708 5MM PEEK MULTIFUNCTION HANDLE ENDOSCOPE AND/OR ACCESSORIES KOG STRYKER ENDOSCOPY-SAN JOSE 07092014

Patients

Seq Age Sex Outcome Treatment
1