FLOGARD
Report
- Report Number
- 1416980-2014-25149
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- July 10, 2014
- Report Date
- July 10, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE FLOGARD INFUSION PUMP WAS SERVICED ON-SITE. A VISUAL INSPECTION, POWER ON SELF-TEST AND AN ALARM LOG REVIEW WERE PERFORMED. DURING THE ALARM LOG REVIEW AND POWER ON SELF-TEST AN F-94 ALARM WAS IDENTIFIED. THE CAUSE OF THE ALARM WAS A DAMAGED BATTERY. THE BATTERY WAS REPLACED AND THE CONFIGURATION SETTINGS WERE RESET TO CORRECT THE REPORTED CONDITION. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A FLOGARD INFUSION PUMP PRESENTED AN F-94 ALARM. IT WAS NOT SPECIFIED WHEN IN THE PROCESS THIS OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 450768 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |