FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3974068
·
Received August 1, 2014
Report
- Report Number
- 3004209178-2014-88581
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- May 9, 2014
- Report Date
- July 2, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
UNIT ALARMED A33 DURING THE PRIME/A33 TEST DUE TO FAULTY FSR. (GOLD) UNIT WAS RECEIVED WITH SCRATCHED LCD WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED ON RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE WINDOW THRU RESERVOIR TUBE AND CRACKED ON BATTERY TUBE THREADS.
Description of Event or Problem · 1
CUSTOMER IS HAVING ISSUES WITH THE INSULIN PUMP. CUSTOMER STATES THAT THE INSULIN PUMP ALARMS DURING PRIME PROCESS. THE BLOOD GLUCOSE READING IS 416 MG/DL. THE INSULIN PUMP WILL BE REPLACED
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 450766 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |