FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3974068 · Received August 1, 2014

Report

Report Number
3004209178-2014-88581
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
May 9, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

UNIT ALARMED A33 DURING THE PRIME/A33 TEST DUE TO FAULTY FSR. (GOLD) UNIT WAS RECEIVED WITH SCRATCHED LCD WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED ON RESERVOIR TUBE LIP, CRACKED RESERVOIR TUBE WINDOW THRU RESERVOIR TUBE AND CRACKED ON BATTERY TUBE THREADS.

Description of Event or Problem · 1

CUSTOMER IS HAVING ISSUES WITH THE INSULIN PUMP. CUSTOMER STATES THAT THE INSULIN PUMP ALARMS DURING PRIME PROCESS. THE BLOOD GLUCOSE READING IS 416 MG/DL. THE INSULIN PUMP WILL BE REPLACED

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
450766 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAP

Patients

Seq Age Sex Outcome Treatment
1 54 YR