FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3974042 · Received August 1, 2014

Report

Report Number
3004209178-2014-89031
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TESTS. THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO A FAULTY GOLD FORCE SENSOR RESISTOR. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. NO MOTOR ERROR ALARM WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP, BROKEN BELT CLIP SLOT AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR AFTER A BOLUS WAS GIVEN. THE BLOOD GLUCOSE READING WAS 167 MG/DL. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. IT WAS FOUND THAT THE CUSTOMER HAD A PACEMAKER, WHICH HAD HAD AN ECHO THE DAY PRIOR WHILE THE CUSTOMER HAD THE INSULIN PUMP ATTACHED. THE CUSTOMER STATED THAT SHE WAS ABLE TO REWIND THE INSULIN PUMP. ADVISED DISCONTINUATION OF THE DEVICE. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
450727 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 75 YR