FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3974017 · Received August 1, 2014

Report

Report Number
3004209178-2014-89023
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH BUTTON ERROR ALARM DUE TO MOISTURE DAMAGE ON KEYPAD TRACES. NO SCROLLING NUMBERS DISPLAY ANOMALY NOTED. THE INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD SCROLLING NUMBERS AND ALARMED BUTTON ERROR DURING A BOLUS ATTEMPT. THE BLOOD GLUCOSE READING WAS 106 MG/DL. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
451005 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1 29 YR