530G INSULIN PUMP

MDR report key: 3974000 · Received August 1, 2014

Report

Report Number
3004209178-2014-89003
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THERE WAS NO BUTTON ERROR ALARM. NO MOISTURE DAMAGE WAS FOUND INSIDE THE INSULIN PUMP. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY SCREEN, CRACKED CASE DISPLAY WINDOW CORNER AND A CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING A BUTTON ERROR ALARM ON THE INSULIN PUMP AFTER KYACKING. HE STATED THAT THEY INSULIN PUMP MAY HAVE BEEN EXPOSED TO WATER. THE BLOOD GLUCOSE READING WAS 125 MG/DL. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
450634 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAB

Patients

Seq Age Sex Outcome Treatment
1 18 YR