FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3973985 · Received August 1, 2014

Report

Report Number
3004209178-2014-88985
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
June 30, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED FOR A MOTOR ERROR DURING THE OCCLUSION TEST AND WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO FAULTY FORCE SENSOR RESISTOR. THE MOTOR PASSED THE MOTOR TEST. THE INSULIN PUMP PASSED THE NO DELIVERY TEST AND NO EXCESSIVE NO DELIVERY ALARMS WERE NOTED. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED RESERVOIR TUBE LIP AND A CRACKED RESERVOIR TUBE WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER CALLED IN TO REPORT THAT THE INSULIN PUMP ALARMED DURING THE PRIME PROCESS. BLOOD GLUCOSE READING AT THE TIME WAS 161 MG/DL. TROUBLESHOOTING WAS PERFORMED. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WILL BE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
450625 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522NAL

Patients

Seq Age Sex Outcome Treatment
1 68 YR