PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-88985
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- June 30, 2014
- Report Date
- July 2, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP ALARMED FOR A MOTOR ERROR DURING THE OCCLUSION TEST AND WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO FAULTY FORCE SENSOR RESISTOR. THE MOTOR PASSED THE MOTOR TEST. THE INSULIN PUMP PASSED THE NO DELIVERY TEST AND NO EXCESSIVE NO DELIVERY ALARMS WERE NOTED. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED RESERVOIR TUBE LIP AND A CRACKED RESERVOIR TUBE WINDOW.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER CALLED IN TO REPORT THAT THE INSULIN PUMP ALARMED DURING THE PRIME PROCESS. BLOOD GLUCOSE READING AT THE TIME WAS 161 MG/DL. TROUBLESHOOTING WAS PERFORMED. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WILL BE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 450625 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-522NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |