FDA Adverse Event Malfunction Summary report: N

3.5MM 90-S SERFAS ENERGY SUCTION PROBE

MDR report key: 3973893 · Received August 1, 2014

Report

Report Number
0002936485-2014-00547
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
June 25, 2014
Report Date
July 4, 2014
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
GEI
PMA / PMN Number
K071859
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE TIP BREAKING CONDITION WAS NOT CONFIRMED SINCE THE UNIT WAS NOT RETURNED. PROBABLE ROOT CAUSES FOR THE REPORTED CONDITION CAN BE ASSOCIATED, BUT ARE NOT LIMITED TO: NON CONFORMING COMPONENT. EVEN THOUGH THE DEVICE HISTORY RECORD OF THE REPORTED UNIT COULD NOT BE REVIEWED SINCE THE LOT NUMBER WAS NOT REPORTED, A POSSIBLE WORKMANSHIP (ASSEMBLY) RELATED CONDITION IS A LESS LIKELY TO BE A CONTRIBUTOR BASED ON THE LINE CONTROLS, NON-CONFORMANCE REPORT AND PREVIOUS COMPLAINT HISTORY RELATED TO THIS PART NUMBER OR CONDITION REPORTED. POOR ASSEMBLY PROCESS. A 30X INSPECTION OF THE PROBE¿S TIP ASSEMBLY IS PERFORMED ON ALL THE PROBES DURING MANUFACTURING. THE ACTIVATION HISTORY COULD NOT BE RETRIEVED SINCE THE UNIT WAS NOT RETURNED FOR EVALUATION. A POSSIBLE WORKMANSHIP (ASSEMBLY) RELATED CONDITION IS A LESS LIKELY POSSIBLE CONTRIBUTOR BASED ON THE LINE CONTROLS AND PREVIOUS COMPLAINT HISTORY. NO PREVIOUS SIMILAR CERAMIC TIP BREAKING CASES HAVE BEEN REPORTED FOR THIS PART NUMBER. MISUSE. THE PROBE IS DELIVERED INSIDE A BLISTER IN WHICH THE UNIT IS SNAPPED IN PLACE, AND THE BLISTER IS PLACED INSIDE A BOX, WHICH PROVIDES PROTECTION TO THE TIP OF THE PROBE. IN THE EVENT THAT THE PROBE IS DELIVERED TO THE CUSTOMER WITH A DAMAGED TIP, THEN THE UNIT MUST BE DISCARDED BEFORE USE. IT IS POSSIBLE THAT THE UNIT WAS USED AS A TOOL FOR THE MECHANICAL DISPLACEMENT OF TISSUE OR BONE, OR USE THE PROBE AS A PRYING OR SCRUBBING TOOL, OR THAT EXCESSIVE FORCE WAS USED WHEN INSERTING OR REMOVING THE PROBE INTO AN OBSTRUCTED PASSAGEWAY, AS THIS MAY RESULT IN DAMAGE TO THE TIP OF THE PROBE. THEREFORE, A POSSIBLE USER RELATED (MISUSE) POSSIBLE CONTRIBUTOR CANNOT BE RULED OUT. IN SUM, THE UNIT WAS NOT RETURNED AND THE FAILURE MODE WAS NOT CONFIRMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS DIFFICULTY RECOVERING THE BROKEN TIP, BUT THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT MEDICAL INTERVENTION, ADVERSE CONSEQUENCES OR SURGICAL DELAY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS DIFFICULTY RECOVERING THE BROKEN TIP, BUT THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT MEDICAL INTERVENTION, ADVERSE CONSEQUENCES OR SURGICAL DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
451047 3.5MM 90-S SERFAS ENERGY SUCTION PROBE ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI STRYKER ENDOSCOPY-SAN JOSE NA

Patients

Seq Age Sex Outcome Treatment
1