FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3973890 · Received August 1, 2014

Report

Report Number
1416980-2014-25140
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 9, 2014
Report Date
July 9, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. THE DEVICE WAS NOT RETURNED; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. INITIATING THERAPY WITHOUT BEING CONNECTED IS A KNOWN CAUSE OF THIS ALARM. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR WHEN AND HOW TO CONNECT TO THE DISPOSABLE SET. IT ALSO INSTRUCTS TO CONNECT THE TRANSFER SET TO THE PATIENT LINE PRIOR TO STARTING THE INITIAL DRAIN. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN SET/LINE) ALARM OCCURRED ON THE HOMECHOICE (HC) DEVICE DURING INITIAL DRAIN. THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE ALARM. DURING TROUBLESHOOTING, IT WAS NOTED THAT THE PATIENT WAS NOT CONNECTED WHEN THERAPY WAS INITIATED. THE PATIENT CONNECTED AFTER INITIATING THERAPY. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP IN CLEARING THE ALARM. THE TSR REVIEWED PROPER PROCEDURES WITH THE HP. THE HP WOULD COMPLETE THERAPY USING ALL NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
451046 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 84 YR HOMECHOICE