ECHELON LINEAR CUTTER
Report
- Report Number
- 3005075853-2014-05436
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- June 6, 2014
- Report Date
- June 10, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K110385
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT ONE PSE60A DEVICE WAS RETURNED WITH NO VISUAL NON-CONFORMANCES AND WITHOUT A CARTRIDGE RELOAD PRESENT. IN ADDITION, A CARTRIDGE PAN WAS FOUND LODGED INSIDE THE CHANNEL. THE PAN WAS REMOVED FROM THE CHANNEL AND THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD. THE DEVICE FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. WHILE NO CONCLUSION COULD BE REACHED AS TO HOW THE PAN BECAME DISLODGED FROM THE RELOAD, IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, THE CARTRIDGE COULD NOT BE LOADED INTO THE DEVICE AT THE 2ND FIRING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 450434 | ECHELON LINEAR CUTTER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |