MEGA SUTURECUT NEEDLE DRIVERINSTRUMENT
Report
- Report Number
- 2955842-2014-04664
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- June 23, 2014
- Report Date
- July 3, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THAT THE GRIP CABLE IS FRAYED AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT STICKS OUT AT THE WRIST. THE DISTAL CLEVIS HUB DOES NOT EXHIBIT ANY WEAR. THE OTHER CABLES AT THE WRIST ARE NOT DAMAGED. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THAT THE DISTAL END OF THE MAIN TUBE HAS VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES ARE NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE FRAYED GRIP CABLE, AND THE ADDITIONAL FINDING OF TUBE ABRASIONS WITH MATERIAL REMOVED, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
THE INSTRUMENT HAS BEEN RETURNED FOR EVALUATION, HOWEVER; FAILURE ANALYSIS INVESTIGATION OF THE RETURNED AFFECTED PART IS NOT COMPLETE. AT THIS TIME, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED AFTER THE EVALUATION HAS BEEN COMPLETED OR IF ADDITIONAL INFORMATION IS RECEIVED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE EXPOSED WIRES COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE WIRES ON THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT WERE SHOWING. THERE WAS NO REPORT THAT ANY PIECE(S) FROM THE INSTRUMENT FELL INTO THE PATIENT AND THERE WAS NO ALLEGATION OF PATIENT HARM OR ADVERSE OUTCOME INVOLVING THE REPORTED INSTRUMENT.