HEARTWARE® VENTRICULAR ASSIST SYSTEM
Report
- Report Number
- 3007042319-2014-00801
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- September 6, 2013
- Report Date
- July 2, 2014
- Manufacturer
- HEARTWARE
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Removal / Correction Number
- Z-1607-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
PRELIMINARY EVALUATION AND TESTING REVEALED THAT THE RETURNED BATTERY CONTAINED A FAULTY CELL. THIS FINDING WAS RE-ASSESSED PER OUR SOP REQUIREMENTS, AND DETERMINED TO BE REPORTABLE. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN THIRTY (30) DAYS OF COMPLETION OF THE INVESTIGATION. THIS IS ONE OF TWO REPORTS (3007042319-2014-00800 AND 3007042319-2014-00801) SUBMITTED FOR DEVICES RELATED TO THE SAME EVENT. EVALUATION IN PROGRESS
THE BATTERIES WERE RETURNED; VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICES IN RELATION TO THE REPORTED EVENT. THOROUGH EXTERNAL VISUAL INSPECTION OF THE BATTERIES REVEALED NO SIGNS OF PHYSICAL DAMAGE OR CONTAMINATION. FUNCTIONAL TESTING REVEALED THAT (B)(4) CONTAINED A FAULTY CELL PAIR WHICH MOST LIKELY CONTRIBUTED TO THE REPORTED EVENT. AN INTERNAL INVESTIGATION (CAPA) HAS BEEN OPENED TO ADDRESS THE ISSUE. THIS IS ONE OF TWO REPORTS (3007042319-2014-00800 AND 3007042319-2014-00801) SUBMITTED FOR DEVICES RELATED TO THE SAME EVENT.
APPROXIMATELY TWO YEARS AND SEVEN MONTHS POST HVAD IMPLANTATION, THE SITE REPORTED THAT THE PATIENT'S BATTERY WAS NOT HOLDING CHARGE. THE BATTERY WAS REMOVED FROM THE PATIENT AND A NEW BATTERY WAS SUPPLIED. NO HARM OR INJURY TO THE PATIENT WAS REPORTED AS A RESULT OF THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 450237 | HEARTWARE® VENTRICULAR ASSIST SYSTEM | CIRCULATORY ASSIST SYSTEM, BATTERY | DSQ | HEARTWARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | (B)(4) - BATTERY| (B)(4) - BATTERY| (B)(4) - BATTERY| (B)(4) - BATTERY |