FDA Adverse Event Malfunction Summary report: N

ENSEAL G2 CURVED JAW

MDR report key: 3973740 · Received August 1, 2014

Report

Report Number
3005075853-2014-05425
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 22, 2014
Report Date
July 24, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K112033
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RECEIVED IN GOOD PHYSICAL CONDITION. THE DEVICE WAS TESTED ON THE GENERATOR AND PASSED ALL FUNCTIONAL TESTING. THE ENERGY OUTPUT DELIVERED FROM THE DEVICE WAS VERIFIED AND THE CUTTING OF THE TEST MEDIA PERFORMED AS EXPECTED. ALL THREE TONES WERE HEARD DURING FUNCTIONAL TESTING (TONE 1 IS HEARD WHEN THE ENERGY ACTIVATION BUTTON IS PRESSED; TONE 2 IS HEARD WHEN TISSUE IMPEDANCE THRESHOLD IS REACHED; AND TONE 3 IS HEARD WHEN THE CYCLE IS COMPLETE). NO ALERT SCREENS WERE DISPLAYED DURING TESTING. NO ISSUES WERE NOTED WITH OPENING AND CLOSING OF THE JAWS. IT IS POSSIBLE THAT EXCESSIVE BODY FLUIDS INFLUENCE THE OPENING AND CLOSING OF THE JAWS. THE ACCUMULATION OF BODY FLUIDS IS A ROUTINE OCCURRENCE DURING SURGICAL PROCEDURES AND DOES NOT NECESSARILY INDICATE A MANUFACTURING DEFECT IN THE DEVICE. THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE, THE JAWS WOULD NOT OPEN (LOT NUMBER L4EF230). REPLACE INSTRUMENT ERROR CODE CAME UP (LOT NUMBER L4EN2V). NO PATIENT CONSEQUENCES REPORTED. PROCEDURE WAS COMPLETED WITH SAME LIKE DEVICE AND PROLONGED BY TEN MINUTES. TWO DEVICES WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
450235 ENSEAL G2 CURVED JAW ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES GEI ETHICON ENDO-SURGERY, LLC. NA L4EF23

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR