HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-25108
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- March 1, 2014
- Report Date
- July 10, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE HIGH DRAIN 105 (NIGHT DRAIN #5) EVENT WAS VERIFIED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED AS NOT ENOUGH EVIDENCE WAS AVAILABLE TO MAKE A DETERMINATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN ERROR 105 (NIGHT DRAIN #5) ALARM WAS IDENTIFIED IN THE LOG (INCLUDES ANY STANDARD MODE THERAPY WHERE THE PATIENT DRAIN VOLUME EXCEEDS 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME). THE ALARM OCCURRED ON (B)(6) 2014 06:35:13.:17:54. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 449813 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |