FDA Adverse Event Malfunction Summary report: N

ETRIO

MDR report key: 3973415 · Received August 1, 2014

Report

Report Number
3005075853-2014-05414
Event Type
Malfunction
Date Received
August 1, 2014
Date of Event
July 24, 2014
Report Date
July 24, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K072177
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RECEIVED IN GOOD PHYSICAL CONDITION. THE DEVICE WAS TESTED ON THE GENERATOR AND PASSED ALL FUNCTIONAL TESTING. THE ENERGY OUTPUT DELIVERED FROM THE DEVICE WAS VERIFIED AND THE CUTTING OF THE TEST MEDIA PERFORMED AS EXPECTED. ALL THREE TONES WERE HEARD DURING FUNCTIONAL TESTING (TONE 1 IS HEARD WHEN THE ENERGY ACTIVATION BUTTON IS PRESSED; TONE 2 IS HEARD WHEN TISSUE IMPEDANCE THRESHOLD IS REACHED; AND TONE 3 IS HEARD WHEN THE CYCLE IS COMPLETE). NO ISSUES WERE NOTED WITH OPENING AND CLOSING OF THE JAWS. THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. THIS REPORT IS NOT INTENDED TO DENY THAT YOU EXPERIENCED A PROBLEM WITH THE DEVICE. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS.

Additional Manufacturer Narrative · 1

(B)(4). NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, AFTER CUTTING TISSUE, THE JAWS WOULD NOT OPEN. NO PATIENT CONSEQUENCES REPORTED. ONE DEVICE WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
450931 ETRIO ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI ETHICON ENDO-SURGERY, LLC. NA K4DW0W

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR