ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-22084
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Report Date
- July 22, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 - DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/13/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX DATA REVEALED MULTIPLE UNEXPLAINED PUMP REBOOTS. DURING TESTING, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED. THE RETURNED BATTERY CAP HAD STRIPPED THREADS, AND WAS UNABLE TO FULLY SECURE ONTO THE PUMP. THE UNEXPLAINED PUMP REBOOTS WERE DUPLICATED WITH THE RETURNED BATTERY CAP. THE PUMP WAS EXERCISED FOR 24 HOURS WITH A TEST BATTERY CAP, AND NO POWER INTERRUPTIONS OR DUPLICATED ALARMS WERE OBSERVED. THE PUMP WAS OPENED AND NO EVIDENCE OF MOISTURE WAS FOUND INSIDE THE PUMP. UNRELATED TO THE INITIAL COMPLAINT, THE DISPLAY SCREEN WAS DISCOLORED.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (DAMAGE) ISSUE. THE REPORTER ALLEGED THAT THE PUMP WAS LOSING POWER INTERMITTENTLY. IT WAS ALLEGED THAT THE BATTERY COMPARTMENT WAS CRACKED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 450982 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |