LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2014-01416
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Date of Event
- April 7, 2014
- Report Date
- May 7, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVAL OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T CHARGE BATTERIES) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY CHARGER/MODEM WAS UNABLE TO DETECT INSERTED BATTERY PACKS. THE CAUSE FOR THE FAILURE WAS ISOLATED TO CONTAMINATION ON THE BATTERY OF THE BEDSIDE PCA BOARD. THE ROOT CAUSE FOR THE CONTAMINATION WAS INGRESS OF AN UNK LIQUID. THE ROOT CAUSE FOR THE SHORTED U13 COMPONENT COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM. THE PATIENT RECEIVED A REPLACEMENT BATTERY CHARGER/MODEM.
THE WIFE OF A (B)(6) MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PATIENT'S BATTERY CHARGER WASN'T CHARGING HIS BATTERIES. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329466 | LIFEVEST WCD 4000 SYSTEM | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |