FDA Adverse Event Malfunction Summary report: N

ENVISION BLOWER KIT

MDR report key: 3973313 · Received June 6, 2014

Report

Report Number
1824206-2014-01729
Event Type
Malfunction
Date Received
June 6, 2014
Date of Event
May 12, 2014
Report Date
May 12, 2014
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE POWER CORD WAS DAMAGED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE POWER CORD TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE UNIT IS SPARKING. THE BED WAS LOCATED IN A PATIENT ROOM AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331529 ENVISION BLOWER KIT BED FLOTATION THERAPY POWERED FNL HILL-ROM INC. 1360

Patients

Seq Age Sex Outcome Treatment
1