FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 3973292 · Received June 6, 2014

Report

Report Number
3006697241-2014-00495
Event Type
Malfunction
Date Received
June 6, 2014
Date of Event
May 9, 2014
Report Date
May 9, 2014
Manufacturer
HILLROM DE MEXICO S DE RL DE CV
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND ONE CASTER NOT HOLDING DUE TO THE BRAKE CASTER SPINDLE NEEDS TO BE ADJUSTED TO TIGHTEN BRAKES. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN ADJUSTED ALL 4 BRAKE CASTER SPINDLE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES ARE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT IN THE BASEMENT. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331456 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILLROM DE MEXICO S DE RL DE CV 8000

Patients

Seq Age Sex Outcome Treatment
1