FDA Adverse Event Malfunction Summary report: N

SINGLE USE INJECTOR

MDR report key: 3973259 · Received June 6, 2014

Report

Report Number
8010047-2014-00324
Event Type
Malfunction
Date Received
June 6, 2014
Date of Event
May 8, 2014
Report Date
May 20, 2014
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
FCG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT PRODUCTS WERE NOT RETURNED FOR EVAL YET. THE EXACT CAUSE OF THE USER'S REPORT COULD NOT BE CONCLUSIVELY DETERMINED. THERE ARE POSSIBILITIES THAT THE INSERTION PORTION OF THE DEVICE WAS DEFORMED AND IT CAUSED THE FAILURE TO EXTEND THE NEEDLE. THE INSTRUCTION MANUAL WARNS USERS THAT "STRAIGHTEN OUT THE INSTRUMENT BEFORE INSPECTING IT. THE INSTRUMENT CAN BE DAMAGED IF IT IS COILED WHILE THE HANDLE IS OPERATED". A SUPPLEMENTAL REPORT WILL BE SUBMITTED, IF ADDITIONAL AND SIGNIFICANT INFO BECOMES AVAILABLE AT A LATER TIME. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION. PLEASE CROSS REF: 8010047-2014-00321/00322/00323.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT DURING UNSPECIFIED PROCEDURE, THE FACILITY CONFIRMED THAT THE NEEDLES WORKED CORRECTLY BUT THEY DID NOT PROJECT OUT OF THE SHEATH WHEN THEY WERE INSERTED INTO THE ENDOSCOPE. THIS HAPPENED ON 4 OCCASIONS. THE DOCTOR COMPLETED THE PROCEDURE WITH A SIMILAR DEVICE. THERE WAS NO REPORT OF PT INJURY REGARDING THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331541 SINGLE USE INJECTOR INJECTION NEEDLE FCG OLYMPUS MEDICAL SYSTEMS CORPORATION NM-400U-0425 42K

Patients

Seq Age Sex Outcome Treatment
1