HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-25048
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- June 24, 2014
- Report Date
- July 9, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE HOMECHOICE DEVICE RECEIVED A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDED FUNCTIONAL AND ELECTRICAL TESTING OF THE DEVICE. THE REPORTED DIFFICULTY OF AN INCREASED INTRA PERITONEAL VOLUME (IIPV) EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS EVENT OCCURRED IN THE THERAPY THAT WAS INITIATED ON (B)(6) 2014 AT 11:22:25. DURING NIGHT DRAIN CYCLE TWO, THE PATIENT'S ULTRAFILTRATION READING WAS 1817ML, INDICATING THE HOME PATIENT (HP) DRAINED 1817ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 3000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 450293 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR |