FDA Adverse Event Malfunction Summary report: N

THUNDERBEAT 5 MM 35 CM FRONT-ACTUATED GRIP

MDR report key: 3973243 · Received June 6, 2014

Report

Report Number
8010047-2014-00307
Event Type
Malfunction
Date Received
June 6, 2014
Date of Event
May 15, 2014
Report Date
May 19, 2014
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICES WERE RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP (OMSC) FOR EVALUATION. THE EVALUATION CONFIRMED THAT THE PTFE PAD PARTIALLY SEPARATED. THERE WERE CONTACT MARKS ON THE SURFACE OF THE PROBE AND JAW. THE MANUFACTURING RECORD WAS REVIEWED WITH NO IRREGULARITIES. BASED ON SIMILAR CASES WITH THE SAME MODEL, IT IS KNOWN THAT BY CONTINUOUSLY ACTIVATING OUTPUT WITHOUT GRASPING ANYTHING IN THE GRASPING SECTION (INCLUDING AFTER THE TISSUE SEPARATED), THE PTFE PAD SEVERELY WEARS, AND THE DISTAL END OF THE PAD SEPARATES FROM THE GRASPING SECTION. CONSIDERING THE EVALUATION RESULT OF THE SUBJECT DEVICE, OMSC CONCLUDED THAT THE REPORTED EVENT OCCURRED SINCE THE USER CONTINUED ACTIVATING OUTPUT FOR AN EXTENDED TIME AFTER THE TISSUE ALREADY CUT. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION. PLEASE CROSS-REFERENCE 8010047-2014-00306, 8010047-2014-00308.

Description of Event or Problem · 1

DURING A LAPAROSCOPIC HEPATECTOMY, 3 SUBJECT DEVICES FAILED. THE PTFE PAD OF 3 SUBJECT DEVICE WAS PARTIALLY SEPARATED AFTER ABOUT 1 HOUR OF USE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. THIS IS A REPORT REGARDING THE SECOND SUBJECT DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331422 THUNDERBEAT 5 MM 35 CM FRONT-ACTUATED GRIP THUNDERBEAT HANDPIECE GEI OLYMPUS MEDICAL SYSTEMS CORPORATION TB-0535FC K3X24

Patients

Seq Age Sex Outcome Treatment
1