THUNDERBEAT 5 MM 35 CM FRONT-ACTUATED GRIP
Report
- Report Number
- 8010047-2014-00307
- Event Type
- Malfunction
- Date Received
- June 6, 2014
- Date of Event
- May 15, 2014
- Report Date
- May 19, 2014
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE SUBJECT DEVICES WERE RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP (OMSC) FOR EVALUATION. THE EVALUATION CONFIRMED THAT THE PTFE PAD PARTIALLY SEPARATED. THERE WERE CONTACT MARKS ON THE SURFACE OF THE PROBE AND JAW. THE MANUFACTURING RECORD WAS REVIEWED WITH NO IRREGULARITIES. BASED ON SIMILAR CASES WITH THE SAME MODEL, IT IS KNOWN THAT BY CONTINUOUSLY ACTIVATING OUTPUT WITHOUT GRASPING ANYTHING IN THE GRASPING SECTION (INCLUDING AFTER THE TISSUE SEPARATED), THE PTFE PAD SEVERELY WEARS, AND THE DISTAL END OF THE PAD SEPARATES FROM THE GRASPING SECTION. CONSIDERING THE EVALUATION RESULT OF THE SUBJECT DEVICE, OMSC CONCLUDED THAT THE REPORTED EVENT OCCURRED SINCE THE USER CONTINUED ACTIVATING OUTPUT FOR AN EXTENDED TIME AFTER THE TISSUE ALREADY CUT. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION. PLEASE CROSS-REFERENCE 8010047-2014-00306, 8010047-2014-00308.
DURING A LAPAROSCOPIC HEPATECTOMY, 3 SUBJECT DEVICES FAILED. THE PTFE PAD OF 3 SUBJECT DEVICE WAS PARTIALLY SEPARATED AFTER ABOUT 1 HOUR OF USE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. THIS IS A REPORT REGARDING THE SECOND SUBJECT DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 331422 | THUNDERBEAT 5 MM 35 CM FRONT-ACTUATED GRIP | THUNDERBEAT HANDPIECE | GEI | OLYMPUS MEDICAL SYSTEMS CORPORATION | TB-0535FC | K3X24 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |