FDA Adverse Event Malfunction Summary report: N

VERSACARE BED

MDR report key: 3973145 · Received June 5, 2014

Report

Report Number
1824206-2014-01724
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
May 8, 2014
Report Date
May 8, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BED SCALE WAS NOT ZEROED, USER ERROR. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012 THROUGH 2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECH ZEROED THE BED SCALE TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THE BED EXIT WILL NOT ARM. THE BED WAS LOCATED IN THE STORAGE ROOM AT THE FACILITY. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
328561 VERSACARE BED A/C POWERED ADJUSTABLE HOSP BED FNL HILL-ROM, INC. 3200

Patients

Seq Age Sex Outcome Treatment
1