FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYS

MDR report key: 3973092 · Received June 5, 2014

Report

Report Number
3008642652-2014-01729
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
May 5, 2014
Report Date
June 4, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CODE 204) WAS CONFIRMED. AS RECEIVED, THERE WAS A SHORT IN THE TRUNK CABLE. THE CAUSE OF THE CODE 204 IS A DISRUPTION IN COMMUNICATION BETWEEN THE ELECTRODE BELT AND MONITOR. THE CAUSE OF THE DISRUPTION IN COMMUNICATION IS THE SHORT. THE ROOT CAUSE OF THE SHORT CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLES AND WIRES. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A SERVICE CODE 204. THE PT WAS ISSUED A REPLACEMENT BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
328550 LIFEVEST WCD 4000 SYS WEARABLE CARDIOVERTER MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 54 YR