FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC BED

MDR report key: 3973086 · Received June 5, 2014

Report

Report Number
1824206-2014-01718
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
May 13, 2014
Report Date
May 13, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE SIDERAIL NOT LATCHING DUE TO THE COVER BENDING. THIS WAS MOST LIKELY DUE TO IT BEING FORCED OPEN OR CLOSED AT SOME POINT, BENDING THE COVER TO THE POINT THAT THE LATCH PHYSICALLY CAUGHT THE COVER. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN BENT BACK THE SIDERAIL METAL COVER TO THE CORRECT SHAPE AND REPAINTED TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE LEFT FOOT SIDERAIL IS NOT LATCHING. THE BED WAS LOCATED AT A HILL-ROM SERVICE CTR NOT IN USE. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
328549 TOTALCARE BARIATRIC BED A/C POWERED ADJUSTABLE HOSP BED FNL HILL-ROM, INC. 1840

Patients

Seq Age Sex Outcome Treatment
1