ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-22072
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Report Date
- July 22, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/13/2014 WITH THE FOLLOWING: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED THAT A CS 078 CALL SERVICE ALARM WAS RECORDED. DURING TESTING, THE PUMP PERFORMED THE REWIND, LOAD, AND PRIME STEPS WITH NO ALARMS OCCURRING. THE PUMP WAS EXERCISED FOR 24 HOURS ON A 1 UNIT PER HOUR BASAL RATE WITH NO ALARMS OCCURRING. THE COMPLAINT THAT THE PUMP WAS EMITTING CALL SERVICE ALARMS WAS OBSERVED IN THE PUMP HISTORY BUT WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 078) ISSUE. IT WAS REPORTED THAT THE PUMP WAS EMITTING MULTIPLE CALL SERVICE ALARMS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 449492 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |