HUT EXT DR FINAL ASSY-REVERSE
Report
- Report Number
- 1518293-2014-00051
- Event Type
- Malfunction
- Date Received
- April 29, 2014
- Date of Event
- April 11, 2014
- Report Date
- April 11, 2014
- Manufacturer
- LIEBEL-FLARSHEIM
- Product Code
- IXR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
MALLINCKRODT TECH SUPPORT (TS) TROUBLESHOT WITH CUSTOMER WHO REPORTED THAT WHEN THEY PRESSED THE FLUORO/RAD PEDALS, NOTHING HAPPENS, NO ERRORS, ONLY A MESSAGE ON THE TABLE DISPLAY THAT STATES PREPARING TO EXPOSE. THEN WHEN THEY LET OFF THE PEDALS, IT STATES EXPOSURE CANCELLED PEDAL RELEASED. TECH SUPPORT FOUND THE CUSTOMER DIDN'T HAVE A PT FILE OPEN AND EXPLAINED THE INFIMED I5 REQUIRES A PT FILE TO BE OPENED IN ORDER TO FLUORO OR PERFORM A DIGITAL SPOT EXPOSURE. CUSTOMER OPENED A TEST PT FILE AND THE SYSTEM ALLOWED FLUORO/DIGITAL SPOTS. SYSTEM IS FULLY FUNCTIONAL, THIS WAS OPERATOR ERROR.
CUSTOMER REPORTS THAT INCIDENT OF NO FLUORO OCCURRED WHEN A PT WAS ON THE TABLE, HOWEVER, THE STAFF WAS STILL ABLE TO COMPLETE THE PROCEDURE USING THE TABLE WITHOUT INCIDENT. NO PT OR PROCEDURAL INFO MADE AVAILABLE. NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 255538 | HUT EXT DR FINAL ASSY-REVERSE | IXR | LIEBEL-FLARSHEIM | HUT EXT DR FINAL A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |