FDA Adverse Event Malfunction Summary report: N

HUT EXT DR FINAL ASSY-REVERSE

MDR report key: 3973044 · Received April 29, 2014

Report

Report Number
1518293-2014-00051
Event Type
Malfunction
Date Received
April 29, 2014
Date of Event
April 11, 2014
Report Date
April 11, 2014
Manufacturer
LIEBEL-FLARSHEIM
Product Code
IXR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MALLINCKRODT TECH SUPPORT (TS) TROUBLESHOT WITH CUSTOMER WHO REPORTED THAT WHEN THEY PRESSED THE FLUORO/RAD PEDALS, NOTHING HAPPENS, NO ERRORS, ONLY A MESSAGE ON THE TABLE DISPLAY THAT STATES PREPARING TO EXPOSE. THEN WHEN THEY LET OFF THE PEDALS, IT STATES EXPOSURE CANCELLED PEDAL RELEASED. TECH SUPPORT FOUND THE CUSTOMER DIDN'T HAVE A PT FILE OPEN AND EXPLAINED THE INFIMED I5 REQUIRES A PT FILE TO BE OPENED IN ORDER TO FLUORO OR PERFORM A DIGITAL SPOT EXPOSURE. CUSTOMER OPENED A TEST PT FILE AND THE SYSTEM ALLOWED FLUORO/DIGITAL SPOTS. SYSTEM IS FULLY FUNCTIONAL, THIS WAS OPERATOR ERROR.

Description of Event or Problem · 1

CUSTOMER REPORTS THAT INCIDENT OF NO FLUORO OCCURRED WHEN A PT WAS ON THE TABLE, HOWEVER, THE STAFF WAS STILL ABLE TO COMPLETE THE PROCEDURE USING THE TABLE WITHOUT INCIDENT. NO PT OR PROCEDURAL INFO MADE AVAILABLE. NO PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
255538 HUT EXT DR FINAL ASSY-REVERSE IXR LIEBEL-FLARSHEIM HUT EXT DR FINAL A NA

Patients

Seq Age Sex Outcome Treatment
1 UNK