HORIZON MEDICAL IMAGING
Report
- Report Number
- 8022257-2014-00005
- Event Type
- Malfunction
- Date Received
- April 29, 2014
- Date of Event
- April 3, 2014
- Report Date
- April 3, 2014
- Manufacturer
- MCKESSON MEDICAL IMAGING COMPANY
- Product Code
- LLZ
- PMA / PMN Number
- K043146
- Removal / Correction Number
- Z-0976-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
WHEN AN ARCHIVING TASK STOPS AND RESTARTS WITHIN THE SAME SECOND, A SOFTWARE DEFECT IN HORIZON MEDICAL IMAGING WILL CREATE A NON-UNIQUE 'JOB ID' RESULTING IN AN INCORRECT ARCHIVE LOCATION FOR A STUDY. AS A RESULT, WHEN A RADIOLOGIST RETRIEVED A STUDY FROM ARCHIVE, INCORRECT IMAGES WERE DISPLAYED FOR THE REQUESTED STUDY. MCKESSON HAS DETERMINED THAT THE IDENTIFIED SOFTWARE DEFECT IS THE ROOT CAUSE FOR THIS PROBLEM. MCKESSON HAS DEVELOPED A SOFTWARE SOLUTION WHICH WILL BE PROVIDED TO POTENTIALLY AFFECTED USERS TO REDUCE THE RISK OF FUTURE RECURRENCES.
HORIZON MEDICAL IMAGING SYSTEM VERSION 11.5.2. PRESENTED AN INCORRECT STUDY (SET OF IMAGES) TO A RADIOLOGIST AT THE REPORTING SITE. THE RADIOLOGIST SELECTED AN ARCHIVED STUDY COMPROMISED OF AN ULTRASOUND ENDOCRINOLOGY IMAGES AND WAS PRESENTED AN ULTRASOUND IMAGES FOR A DIFFERENT PATIENT INSTEAD. NO PATIENT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 255542 | HORIZON MEDICAL IMAGING | PICTURE ARCHIVING AND COMM.SYSTEM | LLZ | MCKESSON MEDICAL IMAGING COMPANY | 11.5.2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |