FDA Adverse Event Malfunction Summary report: N

HUT EXT DR FINAL ASSY-REVERSE

MDR report key: 3973039 · Received April 29, 2014

Report

Report Number
1518293-2014-00046
Event Type
Malfunction
Date Received
April 29, 2014
Date of Event
April 8, 2014
Report Date
April 8, 2014
Manufacturer
LIEBEL-FLARSHEIM CO
Product Code
IXR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT MONITORING FOLLOWED UP WITH THIS CUSTOMER WHO REPORTED THAT THE PROBLEM HAD BEEN RESOLVED. THE STAFF HAD PUSHED THE COMPUTER OVER WHEN REMOVING THE STIRRUPS AND HAD INADVERTENTLY PULLED THE POWER CORD OUT OF THE GIM BOX. BIOMED WAS ABLE TO FIND THIS (POWER CORD OUT) AND AFTER RESEATING THE POWER CORD, ALL FUNCTIONS RESUMED.

Description of Event or Problem · 1

CUSTOMER REPORTS DURING A STONE REMOVAL WITH STENT PLACEMENT THE FLUORO FAILED WITH A GIM COMMUNICATION ERROR. THE PT WAS A (B)(6) YEAR OLD FEMALE. THE PHYSICIAN WAS ABLE TO LOCATE THE STONE FOR REMOVAL WITH THE FLUORO/IMAGE BEFORE THE FAILURE OCCURRED. HE WAS THEN ABLE TO REMOVE THE STONE WITH THE ENDOSCOPY EQUIPMENT AND PLACE THE STENT. NO FOLLOW UP X-RAY WAS REQUIRED. THE PROCEDURE WAS COMPLETED AND NO PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
255581 HUT EXT DR FINAL ASSY-REVERSE IXR LIEBEL-FLARSHEIM CO HUT EXT DR NA

Patients

Seq Age Sex Outcome Treatment
1 60 YR