FDA Adverse Event Injury Summary report: N

PINNACLE MTL INS NEUT36IDX56OD

MDR report key: 3972935 · Received August 1, 2014

Report

Report Number
1818910-2014-24770
Event Type
Injury
Date Received
August 1, 2014
Date of Event
July 10, 2014
Report Date
August 1, 2014
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
KWA
PMA / PMN Number
K003523
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A LOT SPECIFIC COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE PRODUCT AND LOT CODES REQUIRED WERE NOT PROVIDED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY ==> NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER REPORTED INCIDENT(S) AGAINST THE PROVIDED PRODUCT/LOT COMBINATION(S) SINCE RELEASE FOR DISTRIBUTION. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH , A FOLLOW-UP MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

ALTHOUGH THERE ARE NO PATIENT SPECIFIC ALLEGATIONS FOR THIS PATIENT'S RIGHT HIP, WE ARE REOPTING THE HEAD AND LINER BECAUSE GENERAL LITIGATION ALLEGES METAL ON METAL.

Description of Event or Problem · 1

PPF ALLEGES LOOSENING OF CUP AND STEM AND ELEVATED METAL IONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
451430 PINNACLE MTL INS NEUT36IDX56OD PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS KWA DEPUY ORTHOPAEDICS INC US 2415433

Patients

Seq Age Sex Outcome Treatment
1 74 YR Required Intervention