VASCULAR POSITIONING SYSTEM
Report
- Report Number
- 3006795936-2014-00007
- Event Type
- Malfunction
- Date Received
- April 28, 2014
- Date of Event
- April 22, 2014
- Report Date
- April 22, 2014
- Manufacturer
- VASONOVA INC
- Product Code
- OBJ
- PMA / PMN Number
- K123813
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4).
IT WAS REPORTED THE CATHETER WAS BEING PLACED FOR THE PATIENT IN THEIR ROOM. THEY WERE USING THE BIOSENSOR WITH A BARD 6FR POWER PICC SOLO. AN EXPERIENCED PICC NURSE WAS PLACING THE CATHETER WITH THE USE OF THE VPS AND HAD EXCELLENT DOPPLER FLOW. THE ANTI-GRADE FLOW SIGNATURE WAS SEEN AS WELL AS THE HEAR VALVES WERE CLEARLY AUDIBLE. THE CASE ENDED ON A STEADY BLUE BULLS EYE. A CHEST X-RAY WAS PERFORMED AND REVEALED THE CATHETER TIP WAS IN THE CONTRA-LATERAL INTERNAL JUGULAR. THE PATIENT WAS SENT TO INTERVENTIONAL RADIOLOGY TO HAVE THE CATHETER REPOSITIONED. IT WAS NOTED THEY WERE USING A VPS7220B FROM LOT 19F13L0037. THEY ALSO NOTED THEY HAVE DONE APPROXIMATELY 30 TO 40 CASES AND HAVE HAD GREAT SUCCESS BLUE BULLS EYES AND NON-BLUE BULLS EYE CASES. THERE WAS NO DELAY IN THERAPY AND NO PATIENT DEATH OR COMPLICATIONS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 254199 | VASCULAR POSITIONING SYSTEM | CATHETER, ULTRASOUND, INTRAVASCULAR | OBJ | VASONOVA INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | VPS7220B BIOSENSOR, BARD 6 FR POWER PICC SOLO |