PK DISSECTING FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-04657
- Event Type
- Malfunction
- Date Received
- August 1, 2014
- Date of Event
- July 23, 2014
- Report Date
- July 23, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THE INSTRUMENT'S PITCH CABLE TO BE FRAYED AT THE DISTAL IDLER. THERE WAS NO DAMAGE FOUND ON THE CLEVIS. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THE INSTRUMENT'S DISTAL PULLEY HAD SCRATCH MARKS/ABRASIONS. THERE WERE SCRATCHES ON THE SURFACE OF THE DISTAL PULLEY. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE FRAYED CABLE IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT IN CENTRAL PROCESSING, A BROKEN CABLE WAS IDENTIFIED ON THE PK DISSECTING FORCEPS INSTRUMENT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 450947 | PK DISSECTING FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | N10140227 427 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |