FDA Adverse Event
Malfunction
Summary report: N
V200 VENTILATOR
MDR report key: 3972856
·
Received July 8, 2014
Report
- Report Number
- 2031642-2014-00636
- Event Type
- Malfunction
- Date Received
- July 8, 2014
- Report Date
- June 15, 2014
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- CBK
- PMA / PMN Number
- K102054
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
THE INTERNATIONAL CUSTOMER REPORTED THE VENTILATOR CANNOT WORK ON AC POWER DURING NORMAL VENTILATION OPERATION. THE CUSTOMER REPORTED THE DEVICE WAS NOT IN USE ON A PATIENT THEREFORE THERE WAS NO PATIENT INVOLVEMENT OR HARM. THE MANUFACTURERS FIELD SERVICE ENGINEER CONFIRMED THE REPORTED PROBLEM. THE MANUFACTURERS FIELD SERVICE ENGINEER EVALUATED THE DEVICE AND REPLACED THE POWER SUPPLY TO ADDRESS THE REPORTED PROBLEM. THE DEVICE PASSED ALL MANUFACTURER REQUIRED TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 397869 | V200 VENTILATOR | VENTILATOR, CONTINUOUS | CBK | RESPIRONICS CALIFORNIA, INC. | V200 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |